Temperature sensors with unified output TSPU-303p TSMU-303p

3755

Temperature sensors with a unified output TSPU-303p, TSMU-303p allow you to connect temperature sensors to any devices with 4..20 mA, 0..10 V or RS-485 inputs.

Temperature sensors with a unified output TSPU-303p, TSMU-303p or Universal temperature transducers (PTU) are designed to measure temperature and convert the measured value into a signal of 4..20 mA, 0..10 V or RS-485... Temperature sensors 303p are used to measure the temperature of the environment, liquid, gaseous and bulk media in various industries where the outside temperature does not exceed 85 ° C.

Structurally, temperature sensors with a unified output TSPU-303p, TSMU-303p is a sensor of the type 303п with built-in conversion board to 4..20 mA, 0..10 V signal or interface RS-485 with transfer protocol  ModBus RTU.

Universal temperature transducers are designed for:

  • conversion of the resistance value into a current signal (4..20 mA) proportional to the temperature value;
  • conversion of the resistance value into a voltage signal (0..10 V) proportional to the temperature value;
  • measurement and transmission of the measured temperature value via the interface RS-485 (protocol ModBus RTU).

Main characteristics of temperature sensors with unified output TSPU-303p, TSMU-303p

  • Measured temperature range - from -40 to 270 ° С;
  • The minimum length of the mounting part is 80 mm;
  • Power supply from a constant voltage source with a nominal value of 24 V;
  • Converters into a voltage signal (0..10 V are manufactured only in the Z-65 housing);
  • Converters to RS-485 signal are manufactured in G302 housing;
  • The tolerance classes of the sensor installed in the temperature sensor are 1 / 3B (for RTD resistance thermocouples), A, B, C (for ТСМ and ТСП), 1 and 2 (for ТХА, ТХК, ТЖК thermocouples);
  • Thermal converters design - non-separable;
  • Insulation resistance - not less than 100 megohms.

Temperature sensors with a unified output signal can be manufactured with a mounting length from 80 to 3000 mm and a diameter from 4 to 16 mm.

For places of installation where the ambient temperature exceeds 85 ° C, temperature sensors with a unified output signal with a terminal head made of high-temperature press material can be used, for example, type 001 (temperature up to 270 ° С) or duralumin (temperature up to 350 ° С). In this case, a converter board in a separate housing must be used.

Additional Information

Mounting part length L, mm

, , , , , , ,

Supply voltage, V

Accuracy class

,

Output type

, ,

Відгуки

Відгуків немає, поки що.

Send review
REGMIK, NVF, PP
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.